Kaye LTR -90 Operating Manual 9
Chapter 3. Quick Start
Chapter 3. Quick Start
3.1 Unpack and Inspect
Unpack the instrument carefully and examine it for any damage that could have occurred during shipment. If there is
shipping damage, notify Amphenol Advanced Sensors Customer Service and the carrier immediately. Table 2 lists the
equipment and the accessories that comes with the Calibrator. Verify that all the equipment and accessories in Table 6
are in the box.
3.2 Placement
Put the Calibrator on a clean, flat surface. Make sure the Calibrator is 150 mm (6 inches) away from all objects. For
best results, choose a location to set up the product where room temperature changes are minimal.
WARNING!
To prevent possible fire or personal injury:
Do not operate the LTR -90 in orientations other than upright. A fire hazard can be made if the
LTR -90 is put on its side.
Do not remove Inserts when the product shows temperatures more than 50 °C.
Do not operate near flammable materials.
Do not touch the well access surface of the Product.
Do not turn off the LTR -90 when the temperature is above 100 °C. Set a SETPOINT temperature
below 100 °C and let the LTR -90 cool.
Table 2: Parts and Accessories
Name Quantity
LTR -90 1
2-meter (6-foot) Mains Power Cord 1
Certificate of Calibration 1
Insert Removal Tool 1
Well Insulator Cap 1
Temperature Reference to Communication Interface cable 1
Four extruded sleeves (for Thermocouples) 1
Calibration block (with 6 wells) 1