:INIT (:CALibration:PROTected:INITiate)
Purpose
To initiate comprehensive and factory calibration procedures.
Format
:cal:prot:init
Parameter
None
Description
The :INIT command enables Model 2000 calibration when performing these
procedures over the bus. This command must be sent to the unit after sending
the :CODE command, but before sending any other DC, AC, or manufactur-
ing calibration command.
Note
The :INIT command should be sent only once before performing either com-
prehensive, DC, AC, or factory calibration. Do not send :INIT before each
calibration step.
Example
:CAL:PROT:INIT Initiate calibration.
:LOCK
(:CALibration:PROTected:LOCK)
Purpose
To lock out comprehensive or manufacturing calibration.
Format
:cal:prot:lock
Parameter
None
Description
The :LOCK command allows you to lock out both comprehensive and man-
ufacturing calibration after completing those procedures. Thus, :LOCK per-
foms the opposite of enabling calibration with the :CODE command.
Note
To unlock comprehensive calibration, send the :CODE command. To unlock
manufacturing calibration, hold in the OPEN key while turning on the power.
Example
:CAL:PROT:LOCK Lock out calibration.
:LOCK?
(:CALibration:PROTected:LOCK?)
Purpose
To read comprehensive calibration lock status.
Format
:cal:prot:lock?
Response
0 Comprehensive calibration locked.
1 Comprehensive calibration unlocked.
Description
The :LOCK? query requests status from the Model 2000 on calibration
locked/unlocked state. Calibration must be enabled by sending the :CODE
command before calibration can be performed.
Example
:CAL:PROT:LOCK? Request cal lock state.
Calibration Command Reference 3-7