Model 2750 Multimeter/Switch System Service Manual Calibration 2-17
Model 2750 Calibration
DC current calibration
After the 1M resistance point has been calibrated, follow these steps for DC current
calibration:
1. Connect the calibrator to the AMPS and INPUT LO terminals of the Model 2750
as shown in Figure 2-3.
2. Perform the calibration steps listed in Table 2-10. For each step:
Set the calibrator to the indicated current, and make sure the unit is in operate.
(Use the recommended current if possible.)
Send the indicated programming command. (Change the current parameter if
you are using a different calibration current.)
Wait until the Model 2750 completes each step before continuing.
NOTE If you are performing DC-only calibration, proceed to “Programming calibra-
tion dates.”
Tab le 2- 10
DC current calibration programming steps
Calibration
Step
Calibrator
Current Calibration Command* Parameter Range
10mA
100mA
1A
10.00000mA
100.00000mA
1.000000A
:CAL:PROT:DC:STEP10 10E-3
:CAL:PROT:DC:STEP11 100E-3
:CAL:PROT:DC:STEP12 1
9E-3 to 11E-3
90E-3 to 110E-3
0.9 to 1.1
*Change parameter if using different current.