Model 2750 Multimeter/Switch System Service Manual Calibration 2-17
Model 2750 Calibration
DC current calibration
After the 1M resistance point has been calibrated, follow these steps for DC current
calibration:
1. Connect the calibrator to the AMPS and INPUT LO terminals of the Model 2750
as shown in Figure 2-3.
2. Perform the calibration steps listed in Table 2-10. For each step:
Set the calibrator to the indicated current, and make sure the unit is in operate.
(Use the recommended current if possible.)
Send the indicated programming command. (Change the current parameter if
you are using a different calibration current.)
Wait until the Model 2750 completes each step before continuing.
NOTE If you are performing DC-only calibration, proceed to “Programming calibra-
tion dates.”
Tab le 2- 10
DC current calibration programming steps
Calibration
Step
Calibrator
Current Calibration Command* Parameter Range
10mA
100mA
1A
10.00000mA
100.00000mA
1.000000A
:CAL:PROT:DC:STEP10 10E-3
:CAL:PROT:DC:STEP11 100E-3
:CAL:PROT:DC:STEP12 1
9E-3 to 11E-3
90E-3 to 110E-3
0.9 to 1.1
*Change parameter if using different current.

Table of Contents

Other manuals for Keithley 2750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2750 and is the answer not in the manual?

Keithley 2750 Specifications

General IconGeneral
InterfaceGPIB, RS-232, Ethernet
Switching SpeedUp to 500 channels/second
Maximum Voltage300V
Operating Temperature0°C to 50°C
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Temperature
Resolution6.5 digits
Current1A
Maximum Current1A

Related product manuals