Section 2: Performance verification Model DMM6500 -Digit Multimeter with Scanning Calibration and Adjustment Manual
2-38 DMM6500-905-01 Rev. A/July 2018
Figure 19: Connections for DC current verification 10 A range
Apply a relative offset to the DMM6500:
1. On the calibrator, select the OPR/STBY key. Ensure that the front panel displays STANDBY.
2. Verify the DMM6500 zero reading.
3. Verify the DC current verification points and reference DMM readings from the following table.
Verify DC current 10 A range
Description
Nominal input
Typical measured
input (A)
Lower limit
Upper limit
UUT Zero
STANDBY
-2.750E-03
2.750E-03
System zero
0.00E+00
4.80E-04
REL
REL
Full scale (+)
10
9.99961
9.9818606E+00
1.0017359E+01
Half scale (+)
5
4.99984
4.9895902E+00
5.0100898E+00
Half scale ()
5
5.00079
5.0110412E+00
4.9905388E+00
Full scale ()
10
10.0016
1.0019352E+01
9.9838476E+00

Table of Contents

Other manuals for Keithley DMM6500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM6500 and is the answer not in the manual?

Keithley DMM6500 Specifications

General IconGeneral
Resolution6.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
AC Current Range10 µA to 10 A
Temperature Range-200 °C to 1820 °C
Display5-inch capacitive touch screen
Resistance Range100 MΩ
Frequency Range3 Hz
Sampling Rate1 MS/s
ConnectivityUSB, LAN, GPIB (optional), TSP-Link
Memory7 million readings

Related product manuals