Figure 19: Connections for DC current verification 10 A range
Apply a relative offset to the DMM6500:
1. On the calibrator, select the OPR/STBY key. Ensure that the front panel displays STANDBY.
2. Verify the DMM6500 zero reading.
3. Verify the DC current verification points and reference DMM readings from the following table.
Verify DC current 10 A range