Section 2: Performance verification Model DMM6500 -Digit Multimeter with Scanning Calibration and Adjustment Manual
2-38 DMM6500-905-01 Rev. A/July 2018
Figure 19: Connections for DC current verification 10 A range
Apply a relative offset to the DMM6500:
1. On the calibrator, select the OPR/STBY key. Ensure that the front panel displays STANDBY.
2. Verify the DMM6500 zero reading.
3. Verify the DC current verification points and reference DMM readings from the following table.
Verify DC current 10 A range
Description
Nominal input
Typical measured
input (A)
Lower limit
Upper limit
UUT Zero
STANDBY
-2.750E-03
2.750E-03
System zero
0.00E+00
4.80E-04
REL
REL
Full scale (+)
10
9.99961
9.9818606E+00
1.0017359E+01
Half scale (+)
5
4.99984
4.9895902E+00
5.0100898E+00
Half scale ()
5
5.00079
5.0110412E+00
4.9905388E+00
Full scale ()
10
10.0016
1.0019352E+01
9.9838476E+00