EasyManua.ls Logo

Keithley DMM6500

Keithley DMM6500
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model DMM6500 -Digit Multimeter with Scanning Calibration and Adjustment Manual Section 3: Adjustment
DMM6500-905-01 Rev. A/July 2018 3-25
For this adjustment, you need a:
50 Ω coaxial cable
Keithley Model 3390 function generator
Prepare your DMM6500 for a frequency adjustment
To prepare the DMM6500 for a frequency adjustment:
1. Connect the Keithley Instruments Model 3390 function generator to the DMM6500 INPUT HI and
LO terminals as shown in the following figure.
2. Use the BNC to banana adapter at the UUT connection.
Figure 31: Connection for a frequency adjustment
3. Set the function generator output impedance to high, amplitude of 5 V
RMS
, waveform to
square wave.
4. Enable output.
Command parameters for a frequency adjustment
When calibrating your DMM6500 for a frequencyxf adjustment, use the following command
parameters.
Send each command parameter twice. First, send the parameter using the setup command.
Second, send the parameter using the execute command.
The example code below uses the first line of code from the table below. Send all command
parameters using this command.
cal.adjust.step.setup("cal_FREQ_1kHz")
cal.adjust.step.execute("cal_FREQ_1kHz")
Command parameter
Calibrator
function
Calibrator
range
Calibrator
value
Calibrator
frequency
cal_FREQ_1kHz
Frequency
22
10
1 kHz

Table of Contents

Other manuals for Keithley DMM6500

Related product manuals