MLS-BA-defsi-0713
8
8
04 Temperature calibration
We recommend to check the temperature value of the MLS50-3 from time to time.
Before the MLS50-3 should have cooled down at least 3 hours after the last heating
phase. Clamp the temperature sensor (e.g. MLS-A03) in the respective support on
the left side. Push the temperature sensor as close as possible to the thermic sensor
of the MLS.
The temperature is mesaured at two points and can be corrected at the two
temperature points.
1. Carrying out with Temperature calibration set MLB-A03
Ö Use the cursor () to select
04 Temp. calibr.“
Setup29.12.04 Setup
Ö Push -button
The temperature calibration is started.
After 14.59 Min the temp. calibration of the 1. point is
finished.
Ö Select the keys and to correct the
temperature value as necessary (e.g.24/22).
Ö In order to take over the temperature values
press -key, than the temp. calibration for the 2.
point is started.
After 14.59 Min the temp. calibration of the 2. point is
finished.
Ö Select the keys and to correct the
temperature value as necessary (e.g.124/120).
Ö Press key to take over the temperature
values.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN MLS 50-3IR160 and is the answer not in the manual?

KERN MLS 50-3IR160 Specifications

General IconGeneral
BrandKERN
ModelMLS 50-3IR160
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals