MLS-BA-defsi-0713
11
8
11.2 Select finishing criterion
Operator Indication
Ö Actuate the -key, use the cursor () to select
the desired parameter (e.g. “Conclusion“)
Ö Actuate the - key, selection “Finishing
criteria“ is opened
The following turn-off criteria are available:
Ö Turn-off when weighing value changes < 1mg
within 120s
Ö Turn-off when weighing value changes < 1mg
within 20s
Ö Turn-off when weighing value changes
< 1mg within 240s
Ö Turn-off when weighing value changes
< 1mg in freely defined intervals
Ö Turn-off when weighing value changes
Percent range between 0,001% and 9,999%
Ö Turn-off when weighing value changes < 1mg
Ö Turn-off drying process manually
Ö Turn-off drying process according to time (30
seconds)
Select and save settings:
Use the , keys and (Enter) to select the
desired conclusion criteria.