MLS-BA-defsi-0713
94
8.1.2 P2 GLP (Good laboratory practice)
Quality assurance systems require printouts of weighing results as well as of correct
adjustment of the balance stating date and time and balance identification. The
easiest way is to have a printer connected.
Definition of a standard log:
Operator Indication
Ö Call up menu item “P2 GLP“ (see chapter 7.1)
Ö Push -button
P1 alibration
P2 GLP
P3 D
P5 RS-232
P6
P7
C
ate/Time
P4 Redout
Printouts
Globals
Setup29.12.04
01 User
Ö Use the cursor () to select menu item “01 User“
Ö Push -button
Ö Enter user name or number (max. 8 characters)
using the arrow keys (see chapter 7.1); the
respective active position blinks
Setup
P2
01 User
02 Pr
03 Time print
04 Date print
05 User print
oject
ojekt
Last Cal
06 P print
07 Id print
08 Print
WX/235
0
1
Nowak
0
1
1
0
off
on
off
on
on
off
31.01.05
Ö Confirm by pressing the -key
02 Project
Ö Use the cursor () to select “02 Project “
Ö Push -button
Setup
P2
01 User
02 Pr
03 Time print
04 Date print
05 User print
oject
ojekt
Last Cal
06 P print
07 Id print
08 Print
WX/235
0
1
Nowak
0
1
1
0
off
on
off
on
on
off
31.01.05
Ö Enter project name or number (max. 8 characters)
using the arrow keys (see chapter 7.1); the
respective active position blinks
Ö Confirm with -key
All other settings (03 – 08) are carried out in the same
way.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN MLS 50-3IR160 and is the answer not in the manual?

KERN MLS 50-3IR160 Specifications

General IconGeneral
BrandKERN
ModelMLS 50-3IR160
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals