6 Unpacking, installation and commissioning
6.1 Installation Site, Location of Use
The balance is constructed in such a way that reliable weighing results can be
achieved under normal application conditions.
By selecting the correct location for your balance, you will be able to work quickly and
precisely.
Therefore please observe the following at the place of installation:
- Place the balance on a firm, level surface;
- Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
- Protect the balance against direct draughts due to open windows and doors;
- Avoid jarring during weighing;
- Protect the balance against high humidity, vapours and dust;
- Do not expose the device to extreme dampness for longer periods of time.
Inadmissible bedewing (condensation of air moisture on the device) can occur
if a cold device is taken into a significantly warmer environment. In this case,
please acclimatise the device for approx. 2 hours at room temperature after it
has been disconnected from the mains.
- Avoid static charging of the material to be weighed, weighing container and
windshield.
Major display deviations (incorrect weighing results) are possible if electromagnetic
fields occur as well as due to static charging, currents and instable power supply. It is
then necessary to change the location.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN MLS 50-3IR160 and is the answer not in the manual?

KERN MLS 50-3IR160 Specifications

General IconGeneral
BrandKERN
ModelMLS 50-3IR160
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals