9.3 Select turn-off criterion
The user can determine which turn-off criteria are to be used
Operator Indication
Ö Actuate -key, use the cursor (►) to
select the desired setting.
The following turn-off criteria are available:
Ö Turn-off when weighing value changes <
1mg within 120s
Ö Turn-off when weighing value changes
< 1mg within 240s
Ö Turn-off when weighing value changes
< 1mg in freely defined intervals
Ö Turn-off when weighing value changes
Percent range between 0,001% and 9,999%
Ö Turn-off when weighing value changes
< 1mg in freely defined intervals
Ö Turn-off drying process manually
Ö Turn-off drying process time-controlled
Push -button
Actuate -key twice now
Use the -key to save the setting; with the -key the old setting remains valid.