14.1 Remote commands
Commands function
R CR LF
Reset to factory setting – reset
PC CR LF
Command to call up the values of the balance
S CR LF
Measuring value stable
SI CR LF
Measuring value instable
SU CR LF
Last stable measuring value
SUI CR LF
Current measuring value
Z CR LF
Zeroing stable value
ZI CR LF
Zeroing instable value
T CR LF
Taring stable value
TI CR LF
Taring instable value
C0 CR LF
Cancel continuous data output
C1 CR LF
Start continuous data output
NB CR LF
Serial no. of balance
FS CR LF
Weighing range max.
RV CR LF
Software issue
PD CR LF
Date display
PD CR LF
Time display
PMCR LF
Mode of operation
PS CR LF
Balance parameters are printed out
B CR LF
Keytone
ER CR LF
Call up error message
DH CR LF
Display check status bar
CH CR LF
Delete display check status bar
DF CR LF
Display check capacity display
CF CR LF
Display check capacity display
CL CR LF
Start internal adjustment from outside
KL CR LF
Lock keyboard
KU CR LF
Unlock keyboard
E0 CR LF
Keytone off
E1 CR LF
Keytone on
O0 CR LF
Balance off
O1 CR LF
Balance on
A0 CR LF
Auto zero off
A1 CR LF
Auto zero on
PP n CR LF
No. of the user defined data output