PES/PEJ-BA-e-1316
28
6.10.1. Seals and unlocking switch
After verification the balance is sealed at the indicated positions.
Verification of the balance is invalid without the "seal".
Access to the unlocking switch by removing the seal (verification will be invalid!) and
the rubber plug (see drawing).
Position of unlocking switch
Status
forwards
Balance is unlocked for the adjustment
process, adjustment will be possible
backwards
Verification position - Adjustment locked
Position:
Unlocking switch
Seal
Seal