12. Weighing with tolerance range
12.1. General Information
This balance can be used as dispensing as well as sorting balance; the respective
lower tolerance limit as well as upper tolerance limit can be programmed. An acous-
tic signal supports portioning, dispensing or sorting.
In the menu, (see chapter 7) activate the tolerance weighing function:
[ ]
or the combination tolerance weighing/adding (tolerance control on the re-
spective poured quantity):
[ ]
Entering limits is possible for the following functions:
Weighing
Parts counting
Percent determination
Weighing with freely programmable weighing unit
There are two different ways to carry out evaluation of limits:
1. Evaluation of absolute values [ ]:
An exact reference value (e.g. 1 kg) is set.
2. Evaluation with difference values [ ]:
An upper limit and a lower limit for a reference value are set.
Example:
Reference value
Lower limit
Upper limit
Poured quantity
1,000.0 g
970.0 g
1,050.0 g
Evaluation of absolute
values
1,000.0 g
970.0 g
1,050.0 g
Evaluation with differ-
ence values
1,000.0 g
-30.0 g
50.0 g