6 Remote Operations
72 Keysight 11713D/E Operating and Service Manual
CONFigure[:BANK1]:TTL/CONFigure:BANK2:TTL
Turns TTL ON or OFF for each bank. The query commands return supply voltage status
for each bank.
CONFigure[:BANK1]:TTL <bool>
CONFigure:BANK2:TTL <bool>
Examples
To turn TTL ON for bank 1:
CONF:BANK1:TTL 1
To turn TTL OFF for bank 1:
CONF:BANK1:TTL 0
To query the set supply voltage status for bank 1 and bank 2:
CONFigure[:BANK1]:TTL?
CONFigure:BANK2:TTL?
Related Commands
:CONFigure[:BANK1]:X/:CONFigure[:BANK1]:Y
:CONFigure:BANK2:X/:CONFigure:BANK2:Y
CONFigure[:BANK1]/CONFigure:BANK2
CONFigure[:BANK1]:PULSe:DRIVe
CONFigure:BANK2:PULSe:DRIVe
CONFigure[:BANK1]:PULSe:WIDTh
CONFigure:BANK2:PULSe:WIDTh
ATTenuator[:BANK1]:X/ATTenuator[:BANK1]:Y
ATTenuator:BANK2:X/ATTenuator:BANK2:Y
CONFigure[:BANK1]:PULSe:DRIVe
CONFigure:BANK2:PULSe:DRIVe
Sets or returns the instrument’s pulse drive operating mode: 0 for standard mode and 1
for pulse mode.
CONFigure[:BANK1]:PULSe:DRIVe <bool>
CONFigure:BANK2:PULSe:DRIVe <bool>
Query? Yes
Set? Yes
Set/Return Parameters Boolean Data: 0|1|OFF|ON
Minimum n/a
Maximum n/a
Units n/a