Remote Operations 6
Keysight 11713D/E Operating and Service Manual 73
Examples
To set the pulse drive operating mode to 1 (pulse mode):
CONF:PULS:DRIV 1
To set the pulse drive operating mode to 0 (standard mode):
CONF:PULS:DRIV 0
To query the set pulse drive operating mode:
CONFigure[:BANK1]:PULS:DRIV?
CONFigure:BANK2:PULS:DRIV?
Related Commands
:CONFigure[:BANK1]:X/:CONFigure[:BANK1]:Y
:CONFigure:BANK2:X/:CONFigure:BANK2:Y
CONFigure[:BANK1]/CONFigure:BANK2
CONFigure[:BANK1]:TTL/CONFigure:BANK2:TTL
CONFigure[:BANK1]:PULSe:WIDTh
CONFigure:BANK2:PULSe:WIDTh
ATTenuator[:BANK1]:X/ATTenuator[:BANK1]:Y
ATTenuator:BANK2:X/ATTenuator:BANK2:Y
CONFigure[:BANK1]:PULSe:WIDTh
CONFigure:BANK2:PULSe:WIDTh
Sets or returns the instrument’s pulse drive width.
CONFigure[:BANK1]:PULSe:WIDTh <NRf+>
CONFigure:BANK2:PULSe:WIDTh <NRf+>
Query? Yes
Set? Yes
Set/Return Parameters Boolean Data: 0|1|OFF|ON
Minimum n/a
Maximum n/a
Units n/a
Query? Yes
Set? Yes
Set/Return Parameters
Numeric representation format: Integer (for
example, 8)
Minimum 1
Maximum 50
Units milliseconds