FREQuency Subsystem
Command/Query Description
[SOURce:]FREQuency[:CW] <value>|MINimum|MAXimum[,<lower_
limit>,<upper_limit>]
[SOURce:]FREQuency[:CW]? [MINimum|MAXimum]
[SOURce:]FREQuency[:IMMediate] <value>|MINimum|MAXimum
[,<lower_limit>,<upper_limit>]
[SOURce:]FREQuency[:IMMediate]? [MINimum|MAXimum]
Sets the immediate AC frequency and optionally the fre-
quency soft limits.
[SOURce:]FREQuency:LIMit:LOWer <value>|MINimum|MAXimum
[SOURce:]FREQuency:LIMit:LOWer? [MINimum|MAXimum]
[SOURce:]FREQuency:LIMit:UPPer <value>|MINimum|MAXimum
[SOURce:]FREQuency:LIMit:UPPer? [MINimum|MAXimum]
Sets the lower or upper AC frequency soft limit.
[SOURce:]FREQuency:LIMit[:STATe] ON|1|OFF|0
[SOURce:]FREQuency:LIMit[:STATe]?
Enables or disables frequency soft limits.
[SOURce:]FREQuency:MODE FIXed|STEP
[SOURce:]FREQuency:MODE?
Sets the transition mode for frequency settings.
[SOURce:]FREQuency:TRIGgered <value>|MINimum|MAXimum
[SOURce:]FREQuency:TRIGgered? [MINimum|MAXimum]
Sets the triggered AC output frequency when frequency
mode is STEP.
HCOPySubsystem
Command/Query Description
HCOPy:SDUMp:DATA? Returns the display image in BMP format.
Keysight AC6800B Series Operating and Service Guide 109