[SOURce:]CURRent:PROTection:STATe ON|1|OFF|0
[SOURce:]CURRent:PROTection:STATe?
Enables or disables current protection.
Parameter Typical Return
ON|1|OFF|0 0(OFF)or1(ON)
Enable limit control operation: CURR:PROT:STATe OFF
ON (TRIP): Turns OUTPUT off and activates the alarm for overload conditions longer than three
seconds.
OFF(LIMIT CONTROL): Decreases output voltage to keep current at or below the limit when an
overload occurs.
To clear an overcurrent condition, remove the condition's cause and send OUTPut:PROTection:CLEar.
If overcurrent protection is enabled and the output goes into current limit, the output is disabled and
the Questionable Condition status register OCP bit is set.
The *RST command sets this parameter to ON.