VOLTage:RANGe 155 Select the 155 V range.
VOLTage:LIMit:UPPer MAX Sets upper limit to maximum, 315 V
VOLTage:LIMit:LOWer MIN Sets lower limit to minimum, 0 V
VOLTage 120 Sets voltage to 120 V
FREQuency:LIMit:UPPer MAX Sets upper limit to maximum, 500 Hz
FREQuency:LIMit:LOWer MIN Sets lower limit to minimum, 40 Hz
FREQuency 70 Sets frequency to 70 Hz
The commands above set the limits to accommodate the widest possible VOLTage and FREQuency
settings within the given range. On the 155-V voltage range, the VOLTage setting cannot exceed
157.5 V.
In AC+DC mode, the voltage can be specified only when the AC and DC voltage settings are within the
voltage limit range and the peak of the AC+DC waveform is between -445 and 445 V.
Keysight AC6800B Series Operating and Service Guide 77
2îîUser Information