4- 16 Keysight B1500A User’s Guide, Edition 14
Specifications
Hardware
Table 4-9 MCSMU voltage range, resolution, and accuracy
Table 4-10 MCSMU current range, resolution, and accuracy
Range
Force
resolution
Measurement
resolution
Force accuracy
±(%+mV)
a
a. ±(% of output/measured value + fixed offset + proportional offset), Io is the output current in A.
Measurement accuracy
±(%+mV+mV)
a
Maximum
current
±0.2 V 200 nV 200 nV ±(0.06+0.14) ±(0.06+0.14+Io0.05) 1 A
±2 V 2 V2 V ±(0.06+0.6) ±(0.06+0.6+Io0.5) 1 A
±20 V 20 V20 V ±(0.06+3) ±(0.06+3+Io5) 1 A
±40 V
b
b. Maximum output voltage is 30 V.
40 V40 V ±(0.06+3) ±(0.06+3+Io10) 1 A
Range
Force
resolution
Measurement
resolution
Force accuracy ±(%+A+A)
a
a. ±(% of output/measured value + fixed offset + proportional offset), Vo is the output voltage in V.
Measurement accuracy
±(%+A+A)
a
Maximum
voltage
±10 A 10 pA 10 pA ±(0.06+2E-9+Vo1E-10) ±(0.06+2E-9+Vo1E-10) 30 V
±100 A 100 pA 100 pA ±(0.06+2E-8+VoE-9) ±(0.06+2E-8+VoE-9) 30 V
±1 mA 1 nA 1 nA ±(0.06+2E-7+Vo1E-8) ±(0.06+2E-7+Vo1E-8) 30 V
±10 mA 10 nA 10 nA ±(0.06+2E-6+VoE-7) ±(0.06+2E-6+VoE-7) 30 V
±100 mA 100 nA 100 nA ±(0.06+2E-5+VoE-6) ±(0.06+2E-5+VoE-6) 30 V
±1 A
b
b. Pulse mode only. The maximum value of the base current during pulsing is ±50 mA.
1 A1 A ±(0.4+2E-4+VoE-5) ±(0.4+2E-4+VoE-5) 30 V