Keysight B1500A User’s Guide, Ed ition 14 4- 17
Specifications
Hardware
SMU Common Specifications
Output terminal/connection
Dual triaxial connector, Kelvin (remote sensing)
Voltage/current compliance (limiting)
The SMU can limit output voltage or current to prevent damaging the device
under test.
Voltage compliance:
0 V to ± 100 V (MPSMU, HRSMU)
0 V to ± 200 V (HPSMU)
0 V to ± 30 V (MCSMU)
Current compliance:
± 10 fA to ± 100 mA (HRSMU/MPSMU with ASU)
± 100 fA to ± 100 mA (HRSMU)
± 1 pA to ± 100 mA (MPSMU)
± 1 pA to ± 1 A (HPSMU)
± 10 nA to ± 1 A (MCSMU)
Compliance accuracy:
Same as the current or voltage set accuracy.
Pulsed measurement
Programmable pulse width, period, and delay:
For HPSMU, MPSMU, HRSMU
Width: 500 s to 2 s, 100 s resolution
Period: 5 ms to 5 s, 100 s resolution
Period Width+2 ms (when Width 100 ms)
Period  Width+10 ms (when Width > 100 ms)
Delay: 0 s

Table of Contents