SCPI Commands Appendix C
Series E4360 User’s Guide 107
SCPI Command
Description
:PROTection
[:LEVel] <NRf>, (@chanlist)
Sets the over-voltage protection level
:SAS
:VMP <NRf>, (@chanlist)
Sets the voltage at the peak power point of the curve
:VOC <NRf>, (@chanlist)
Sets the open-circuit voltage
:SLIMit
:HIGH <NRf+> | MAXimum [,(@chanlist)]
Sets the high soft limit for programming the output voltage
:TABle
:OFFSet <NRf> [,(@<chanlist>)]
Adds a voltage offset when operating in Table mode
STATus
:OPERation
[:EVENt]? [(@chanlist)]
Returns the value of the operation event register
:CONDition? [(@chanlist)]
Returns the value of the operation condition register
:ENABle <NRf> [,(@chanlist)]
Enables specific bits in the Event register
:NTRansition <NRf> [,(@chanlist)]
Sets the Negative transition filter
:PTRansition <NRf> [, (@chanlist)]
Sets the Positive transition filter
:PRESet
Presets all enable and transition registers to power-on
:QUEStionable
[:EVENt]? [(@chanlist)]
Returns the value of the questionable event register
:CONDition? [(@chanlist)]
Returns the value of the questionable condition register
:ENABle <NRf> [,(@chanlist)]
Enables specific bits in the Event register
:NTRansition <NRf> [,(@chanlist)]
Sets the Negative transition filter
:PTRansition <NRf> [,(@chanlist)]
Sets the Positive transition filter
SYSTem
:CHANnel
[:COUNt]?
Returns the number of output channels in a mainframe
:MODel? [(@chanlist)]
Returns the model number of the selected channel
:OPTion? [(@chanlist)]
Returns the option installed in the selected channel
:SERial? [(@chanlist)]
Returns the serial number of the selected channel
:COMMunicate
:RLSTate LOCal | REMote | RWLock
Specifies the Remote/Local state of the instrument
:ERRor?
Returns the error number and error string
:GROup
:CATalog?
Returns the groups that have been defined
:DEFine [(@chanlist)]
Group multiple channels together to create a single output
:DELete <channel>
Removes the specified channel from a group
:ALL
Ungroups all channels
:PARallel AUTO | DIRect [, (@chanlist)]
Specifies how output modules are connected in parallel
:MMEMory
:INITialize
Initializes the instrument's memory to the factory-default
settings
:SANitize
Implements the sanitizing standard DoD 5220.22-M
:PASSword
:FPANel
:RESet
Resets the front panel lock password to zero
:REBoot
Returns the unit to its power-on state
:VERSion?
Returns the SCPI version number
TRIGger
[:TRANsient]
[:IMMediate] [(@chanlist)
Triggers the output immediately
:SOURce
BUS|DWEL<n>|PIN<n>|TRAN<n>[,(@chanlist)]
Sets the output trigger source