128 Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
10 Triggers
• Digital channel (on mixed-signal oscilloscopes), D0 to the number of digital
channels minus one.
• External — triggers on the rear panel EXT TRIG IN signal.
• Line — triggers at the 50% level of the rising or falling edge of the AC power
source signal.
• WaveGen — triggers at the 50% level of the rising edge of the waveform
generator output signal. (Not available when the DC or Noise waveforms are
selected.)
You can choose a channel that is turned off (not displayed) as the source for the
edge trigger.
The selected trigger source is indicated in the upper-right corner of the display
next to the slope symbol:
• 1 through 4 = analog channels.
• D0 through Dn = digital channels.
• E = External trigger input.
• L = Line trigger.
• W = Waveform generator.
4 Press the Slope softkey and select rising edge, falling edge, alternating edges,
or either edge (depending on the selected source). The selected slope is
displayed in the upper-right corner of the display.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies MSO-X 2022A and is the answer not in the manual?

Keysight Technologies MSO-X 2022A Specifications

General IconGeneral
BrandKeysight Technologies
ModelMSO-X 2022A
CategoryTest Equipment
LanguageEnglish

Related product manuals