Measurements 14
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide 187
Maximum
Maximum is the highest value in the waveform display. The Y cursor shows the
value being measured.
Minimum
Minimum is the lowest value in the waveform display. The Y cursor shows the
value being measured.
Amplitude
The Amplitude of a waveform is the difference between its Top and Base values.
The Y cursors show the values being measured.
Top
The Top of a waveform is the mode (most common value) of the upper part of the
waveform, or if the mode is not well defined, the top is the same as Maximum. The
Y cursor shows the value being measured.
See Also • “To isolate a pulse for Top measurement" on page 187
To isolate a pulse for Top measurement
The following figure shows how to use Zoom mode to isolate a pulse for a Top
measurement.
You may need to change the measurement window setting so that the
measurement is made in the lower, Zoom window. See “Measurement Window
with Zoom Display" on page 200.