Horizontal Controls 2
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide 55
6 Move the Y1 and Y2 cursors to the intersection of the signal and the Y axis.
Again, note the ΔY value.
7 Calculate the phase difference using the formula below.
For example, if the first ΔY value is 1.688 and the second ΔY value is 1.031:
Figure 4 Phase difference measurements, automatic and using cursors
sinθ =
second Δ Y
first Δ Y
=
1.031
1.688
; θ = 37.65 degrees of phase shift
NOTE
Z-Axis Input in XY Display Mode (Blanking)
When you select the XY display mode, the time base is turned off. Channel 1 is the X-axis
input, channel 2 is the Y-axis input, and the rear panel EXT TRIG IN is the Z-axis input. If you
only want to see portions of the Y versus X display, use the Z-axis input. Z-axis turns the trace
on and off (analog oscilloscopes called this Z-axis blanking because it turned the beam on and
off). When Z is low (<1.4 V), Y versus X is displayed; when Z is high (>1.4 V), the trace is turned
off.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies MSO-X 2022A and is the answer not in the manual?

Keysight Technologies MSO-X 2022A Specifications

General IconGeneral
BrandKeysight Technologies
ModelMSO-X 2022A
CategoryTest Equipment
LanguageEnglish

Related product manuals