118 Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
Service and Utility Menus
Utility Service Menu
adjustment will take to complete. Because of this, the recommended
connection type is USB. When selecting a power meter/sensor to use for
this adjustment the following parameters need to be considered:
Frequency Range: 10 MHz to 6 GHz
Power Level: -70 to +20 dBm
While there may be other sensors that meet the required parameters, the
recommended USB power sensors are shown in Table 5-2. When using one
of these USB power sensors it will take approximately 1 hour for them all to
complete.
Be sure to configure and test the communication with the power sensor
prior to executing this routine. This would also include zeroing and
calibrating the power sensor as needed.
If the following error occurs when running this routine, there is a required
adjustment missing. Go to the I/Q Internal Channel Correction Calibrations
section in this chapter and run the “Enhanced Factory Calibration”.
517, Calibration Failure; Internal Channel Correction
Calibration(new)failure. Prerequisite calibration not completed.
Finalize MW Assembly Installation
Not used.
Power Verify
This feature can be used to verify the output power of the instrument. It will
measure multiple frequency points across the frequency range at a few
different power level settings. However, this is not intended for servicing of the
product or performance verification, and any failures of this should not be
considered a failure of the instrument. Refer to Chapter 18, “Performance
Verification and Adjustment Software” for performance verification testing.