FUNCtion Command
[SOURce:]FUNCtion CURRent|VOLTage
[SOURce:]FUNCtion?
Sets the output regulation - voltage priority or current priority. In voltage priority mode the output is
controlled by a constant voltage feedback loop, which maintains the output voltage at its
programmed setting. In current priority mode the output is controlled by a constant current feedback
loop, which maintains the output current at its positive or negative programmed setting.
Parameter Typical Return
CURRent|VOLTage, *RST VOLTage CURR or VOLT
Sets the output regulation to current priority: FUNC CURR
6 SCPI Programming Reference
231 Keysight N6900/N7900 Series Operating and Service Guide

Table of Contents