N6951A/N6971A with Option 301
N6951A/N6971A Test Record Report Number ______ Date__________
Test Description Model Min. Specs Results Max. Specs
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
+ 18 mV
Vout 2 mV
19.992 V
Vout 8 mV
_______
_______
_______
_______
+ 22 mV
Vout + 2 mV
20.008 V
Vout + 8 mV
CV Load Effect: Both 0.75 mV _______ + 0.75 mV
CV Ripple and Noise
peak-to-peak:
rms:
Both
Both
N/A
N/A
_______
_______
+ 9 mV
+ 1 mV
Transient response @ 100 µs: Both 150 mV _______ + 150 mV
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
N6951A
N6971A
N6951A
N6971A
N6951A
N6971A
N6951A
N6971A
8 mA
15 mA
Iout 8 mA
Iout 15 mA
49.972 A
99.945 A
Iout 0.028 A
Iout 0.055 A
_______
_______
_______
_______
_______
_______
_______
_______
+ 8 mA
+ 15 mA
Iout + 8 mA
Iout + 15 mA
50.028 A
100.055 A
Iout + 0.028 A
Iout + 0.055 A
Low Current Measurement Range: N6951A
N6971A
Iout 0.0035 A
Iout 0.007 A
_______
_______
Iout + 0.0035 A
Iout + 0.007 A
CC Load Effect: N6951A
N6971A
3 mA
6 mA
_______
_______
+ 3 mA
+ 6 mA
Current Sink Tests
10% of current rating:
100% of current rating:
N6951A
N6971A
N6951A
N6971A
5.01 A
10.019 A
50.028 A
100.055 A
_______
_______
_______
_______
4.99 A
9.981 A
49.972 A
99.945 A
N6951ASettings N6971ASettings
Voltage Programming & Readback Min:
Voltage Programming & Readback High:
CV Load Effect, CV Ripple and Noise:
Transient Response:
Current Programming & Readback Min:
Current Programming & Readback High:
Low Current Measurement Range:
CC Load Effect:
10% of current rating (Isink):
100% of current rating (Isink):
Voltage priority: 20 mV, 50 A
Voltage priority: 20 V, 50 A
Voltage priority: 20 V, 50 A
Voltage priority: 20 V, 25 A to 50 A
Current priority: 20 V, 0 A
Current priority: 20 V, 50 A
Current priority: 20 V, 5 A
Current priority: 20 V, 50 A
Current priority: 20 V, - 5 A
External supply: 20 V, 5.5 A
Current priority: 20 V, - 50 A
External supply: 20 V, 55 A
Voltage priority: 20 mV, 100 A
Voltage priority: 20 V, 100 A
Voltage priority: 20 V, 100 A
Voltage priority: 20 V, 50 A to 100 A
Current priority: 20 V, 0 A
Current priority: 20 V, 100 A
Current priority: 20 V, 10 A
Current priority: 20 V, 100 A
Current priority: 20 V, - 10 A
External supply: 20 V, 11 A
Current priority: 20 V, - 100 A
External supply: 20 V, 110 A