[SOURce:]LIST:CURRent[:LEVel] <value>{,<value>}
[SOURce:]LIST:CURRent[:LEVel]?
[SOURce:]LIST:VOLTage[:LEVel] <value>,{<value
[SOURce:]LIST:VOLTage[:LEVel]?
Specifies the setting for each list step. Values are specified in either amperes or volts.
Parameter Typical Return
Voltage: 0 to 102% of rating
Current: -10.2% to 102% of rating
Current with power dissipator: up to -102% to 102% of rating
<list value 1>,
<list value 2>,
<list value 3>
Programs a current list. The list contains 3 steps:LIST:CURR 3,2,1
Programs a voltage list. The list contains 3 steps:LIST:VOLT 20,10,5
[SOURce:]LIST:DWELl <value>{,<value>}
[SOURce:]LIST:DWELl?
Specifies the dwell time for each list step. Dwell time is the time that the output will remain at a
specific step. Dwell times can be programmed from 0 through 262.144 seconds with the following
resolution:
Range in seconds Resolution
0 - 0.262144 1 microsecond
0.262144 - 2.62144 10 microseconds
2.62144 - 26.2144 100 microseconds
26.2144 - 262.144 1 millisecond
Parameter Typical Return
0 – 262.144 <list value 1>,
<list value 2>,
<list value 3>
Programs a dwell list. The list contains 3 steps: LIST:DWEL 0.2,0.8,1.6
[SOURce:]LIST:CURRent:POINts?
[SOURce:]LIST:DWELl:POINts?
[SOURce:]LIST:VOLTage:POINts?
[SOURce:]LIST:TOUTput:BOSTep:POINts?
[SOURce:]LIST:TOUTput:EOSTep:POINts?
Returns the number of list points. Points are the same as steps. The queries do not return the point
values.