SYSTem:ERRor?
Reads and clears one error from the error queue.
Parameter Typical Return
(none) <+0,"No error">
Reads and clear first error in error queue: SYST:ERR?
l The front-panel ERR annunciator turns on when one or more errors are currently stored in the
error queue. Error retrieval is first-in-first-out (FIFO), and errors are cleared as you read
them.When you have read all errors from the error queue, the ERR annunciator turns off.
l If more than 20 errors have occurred, the last error stored in the queue (the most recent error) is
replaced with -350,"Error queue overflow". No additional errors are stored until you remove errors
from the queue. If no errors have occurred when you read the error queue, the instrument
responds with +0,"No error".
l The error queue is cleared by the *CLS and when power is cycled. It is not cleared by a *RST.
l Errors have the following format (the error string may contain up to 255 characters).
<error code>,<error string>
For a list of error codes and message strings, see SCPI Error Messages.
SYSTem:LFRequency?
Returns the power-line reference frequency. This determines the integration time used by
SENSe:SWEep:NPLC command.
At power-on, if the line frequency mode is set to Auto, the power supply automatically detects the
power-line frequency (50 Hz, 60 Hz, or 400 Hz) and uses this value to determine the integration time
used. If the detected power line frequency is 400 Hz, the 50 Hz reference value is actually used (a
subharmonic of 400 Hz).
If the auto line detect fails because the line is noisy or out of tolerance, it uses a setting of 60 Hz.
Parameter Typical Return
(none) 50, 60, or 400
Query the power line frequency: SYST:LFR?
SYSTem:LFRequency:MODE AUTO|MAN50|MAN60
SYSTem:LFRequency:MODE?
Specifies automatic or manual line frequency detection.
AUTO specifies automatic detection.
MAN50 specifies a setting of 50 Hz.
MAN60 specifies a setting of 60 Hz. This parameter is saved in non-volatile memory.
6 SCPI Programming Reference
273 Keysight N6900/N7900 Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N6972A with Option 301 and is the answer not in the manual?

Keysight Technologies N6972A with Option 301 Specifications

General IconGeneral
BrandKeysight Technologies
ModelN6972A with Option 301
CategoryPower Supply
LanguageEnglish

Related product manuals