[SOURce:]VOLTage:LIMit[:POSitive][:IMMediate][:AMPLitude] <value>|MIN|MAX
[SOURce:]VOLTage:LIMit[:POSitive][:IMMediate][:AMPLitude]? [MIN|MAX]
Sets the voltage limit when in current priority mode. Units are in volts.
Parameter Typical Return
0.1% to 102% of rating, *RST 1% of rating <voltage limit>
Sets the voltage limitto 20 V: VOLT:LIM 20
[SOURce:]VOLTage:MODE FIXed|STEP|LIST|ARB
[SOURce:]VOLTage:MODE?
Sets the transient mode. This determines what happens to the output voltage when the transient
system is initiated and triggered.
FIXed keeps the output voltage at its immediate value.
STEP steps the output to the triggered level when a trigger occurs.
LIST causes the output to follow the list values when a trigger occurs.
ARB causes the output to follow the arbitrary waveform values when a trigger occurs.
Parameter Typical Return
FIXed|STEP|LIST|ARB, *RST FIXed FIX, STEP, LIST, or ARB
Sets the voltage mode to Step: VOLT:MODE STEP
[SOURce:]VOLTage:PROTection[:LEVel] <value>|MIN|MAX
[SOURce:]VOLTage:PROTection[:LEVel]? [MIN|MAX]
Sets the over-voltage protection level. Units are in volts. If the output voltage exceeds the OVP level,
the output is disabled and the Questionable Condition status register OV bit is set.
Parameter Typical Return
0 to 120% of rating, *RST 120% of rating <protect level>
Sets the over-voltage protection to 24 V: VOLT:PROT 24
l An over-voltage condition can be cleared with the OUTput:PROTection:CLEar command after the
cause of the condition has been removed.
[SOURce:]VOLTage:RESistance[:LEVel][:IMMediate][:AMPLitude] <value>|MIN|MAX
[SOURce:]VOLTage:RESistance[:LEVel][:IMMediate][:AMPLitude]? [MIN|MAX]
Sets the output resistance level in voltage in voltage priority mode. Units are in ohms. Refer to Set the
Output Resistance for the model-specific resistance programming ranges.
Keysight N6900/N7900 Series Operating and Service Guide 282
6 SCPI Programming Reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N6972A with Option 301 and is the answer not in the manual?

Keysight Technologies N6972A with Option 301 Specifications

General IconGeneral
BrandKeysight Technologies
ModelN6972A with Option 301
CategoryPower Supply
LanguageEnglish

Related product manuals