CALibrate Subsystem
Calibrate commands calibrate the instrument.
Read the calibration section before calibrating. Improper calibration reduces
accuracy and reliability.
CALibrate:COUNt?
Returns the number of times the unit has been calibrated. The count is incremented whenever the
calibration (and date) is saved, the administration password is changed or reset, or the firmware is
updated.
Parameter Typical Return
(none) <count>
Return the calibration count: CAL:COUN?
CALibrate:CURRent[:LEVel] <value>
Calibrates the current programming and measurement. The value selects the range to calibrate.
Parameter Typical Return
The maximum current of the output range. (none)
Calibrates the current of the 10 A range: CAL:CURR 10
CALibrate:CURRent:SHARing
Calibrates the Imon signal for paralleled units.
Parameter Typical Return
(none) (none)
Calibrates the current sharing: CAL:CURR:SHAR
CALibrate:CURRent:TC
Calibrates the temperature coefficient.
Parameter Typical Return
(none) (none)
Calibrates the temperature coefficient: CAL:CURR:TC
Keysight RP7900 Series Operating and Service Guide 179
5 SCPI Programming Reference