SENSe Subsystem
Sense commands control the current measurement ranges and window as well as the data
acquisition sequence.
SENSe:AHOur:RESet
SENSe:WHOur:RESet
Resets the amp-hour or watt-hour measurement to zero.
Parameter Typical Return
(none) (none)
Resets the amp-hour measurement: SENS:AHO:RES
Resets the watt-hour measurement: SENS:WHO:RES
SENSe:ELOG:FUNCtion:CURRent 0|OFF|1|ON
SENSe:ELOG:FUNCtion:CURRent?
SENSe:ELOG:FUNCtion:VOLTage 0|OFF|1|ON
SENSe:ELOG:FUNCtion:VOLTage?
Enables or disables the Elog current or voltage measurement function.
Parameter Typical Return
0|OFF|1|ON, *RST OFF 0or1
Enables datalog current measurements: SENS:ELOG:FUNC:CURR ON
SENSe:ELOG:FUNCtion:CURRent:MINMax 0|OFF|1|ON
SENSe:ELOG:FUNCtion:CURRent:MINMax?
SENSe:ELOG:FUNCtion:VOLTage:MINMax 0|OFF|1|ON
SENSe:ELOG:FUNCtion:VOLTage:MINMax?
Enables or disables logging of the minimum and maximum current or voltage values.
Parameter Typical Return
0|OFF|1|ON, *RST OFF 0or1
Enables MIN/MAX logging values: SENS:ELOG:FUNC:VOLT:MINM ON
SENSe:ELOG:PERiod <value>|MIN|MAX
SENSe:ELOG:PERiod? [MIN|MAX]
Sets the integration time of an Elog measurement.
Keysight RP7900 Series Operating and Service Guide 217
5 SCPI Programming Reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies RP7900 Series and is the answer not in the manual?

Keysight Technologies RP7900 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelRP7900 Series
CategoryPower Supply
LanguageEnglish

Related product manuals