Specifications and Characteristics - RP795xA, RP796xA
Specifications - RP795xA, RP796xA
Supplemental Characteristics - RP795xA, RP796xA
Common Characteristics
Output Impedance Graphs
Inductive Load Boundary for Current Priority Mode
Capacitive Load Boundary for Voltage Priority Mode
Voltage Programming Response
Excessive Dynamic Protection
Output Quadrants
Dimensions
Specifications - RP795xA, RP796xA
Unless otherwise noted, specifications are warranted over the ambient temperature range of 0 to
40°C after a 30-minute warm-up period. Specifications apply at the output terminals, with the sense
terminals connected to the output terminals (local sensing).
Specification RP7951A, RP7961A RP7952A, RP7962A RP7953A, RP7963A
DC Ratings
Voltage source:
Current source and sink:
Power:
0 to 500 V
0 to ±20 A
0 to ±5 kW
0 to 500 V
0 to ±40 A
0 to ±10 kW
0 to 950 V
0 to ±20 A
0 to ±10 kW
Output ripple & noise
CV peak-to-peak:
1
CV rms:
2
500 mV
100 mV
500 mV
100 mV
1000 mV
200 mV
Load regulation
Voltage:
Current:
30 mV
9 mA
30 mV
17 mA
60 mV
9 mA
Voltage programming accuracy:
3
0.03% + 60 mV 0.03% + 60 mV 0.03% + 120 mV
Voltage measurement accuracy:
3
0.03% + 80 mV 0.03% + 80 mV 0.03% + 160 mV
Current programming &
measurement accuracy:
3
0.1% + 12 mA 0.1% + 24 mA 0.1% + 12 mA
Transient response
4
Recovery Time:
Settling band:
500 µs
1.25 V
500 µs
1.25 V
500 µs
2.375 V
1
From 20 Hz to 20 MHz (-3dB bandwidth) with resistive load, terminals ungrounded, or either terminal grounded
2
From 20 Hz to 10 MHz (-3dB bandwidth) with resistive load, terminals ungrounded, or either terminal grounded
3
Percent of value + offset; at 25°C ±5°C after a 30 minute warm-up; measurement NPLC=1; valid for 1 year, see Calibration Interval
with a load lead voltagedrop of up to 1V per lead - see Wire Size for additional information about load lead voltagedrop
4
Time to recover to within the settling band following astep change from 50% to 100% of full load at the maximum slew rate
1 Quick Reference
52 Keysight RP7900 Series Operating and Service Guide