Specifications and Characteristics - RP795xA, RP796xA
Specifications - RP795xA, RP796xA
Supplemental Characteristics - RP795xA, RP796xA
Common Characteristics
Output Impedance Graphs
Inductive Load Boundary for Current Priority Mode
Capacitive Load Boundary for Voltage Priority Mode
Voltage Programming Response
Excessive Dynamic Protection
Output Quadrants
Dimensions
Specifications - RP795xA, RP796xA
Unless otherwise noted, specifications are warranted over the ambient temperature range of 0 to
40°C after a 30-minute warm-up period. Specifications apply at the output terminals, with the sense
terminals connected to the output terminals (local sensing).
Specification RP7951A, RP7961A RP7952A, RP7962A RP7953A, RP7963A
DC Ratings
Voltage source:
Current source and sink:
Power:
0 to 500 V
0 to ±20 A
0 to ±5 kW
0 to 500 V
0 to ±40 A
0 to ±10 kW
0 to 950 V
0 to ±20 A
0 to ±10 kW
Output ripple & noise
CV peak-to-peak:
1
CV rms:
2
500 mV
100 mV
500 mV
100 mV
1000 mV
200 mV
Load regulation
Voltage:
Current:
30 mV
9 mA
30 mV
17 mA
60 mV
9 mA
Voltage programming accuracy:
3
0.03% + 60 mV 0.03% + 60 mV 0.03% + 120 mV
Voltage measurement accuracy:
3
0.03% + 80 mV 0.03% + 80 mV 0.03% + 160 mV
Current programming &
measurement accuracy:
3
0.1% + 12 mA 0.1% + 24 mA 0.1% + 12 mA
Transient response
4
Recovery Time:
Settling band:
500 µs
1.25 V
500 µs
1.25 V
500 µs
2.375 V
1
From 20 Hz to 20 MHz (-3dB bandwidth) with resistive load, terminals ungrounded, or either terminal grounded
2
From 20 Hz to 10 MHz (-3dB bandwidth) with resistive load, terminals ungrounded, or either terminal grounded
3
Percent of value + offset; at 25°C ±5°C after a 30 minute warm-up; measurement NPLC=1; valid for 1 year, see Calibration Interval
with a load lead voltagedrop of up to 1V per lead - see Wire Size for additional information about load lead voltagedrop
4
Time to recover to within the settling band following astep change from 50% to 100% of full load at the maximum slew rate
1 Quick Reference
52 Keysight RP7900 Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies RP7900 Series and is the answer not in the manual?

Keysight Technologies RP7900 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelRP7900 Series
CategoryPower Supply
LanguageEnglish

Related product manuals