OUTPut:RELay:LOCK[:STATe] 0|OFF|1|ON
OUTPut:RELay:LOCK[:STATe]?
Enables or disables the locked relay-state of the Keysight SD1000A Safety Disconnect System. When
locked, the output relays of the SDS remain closed and do not change along with the output state of
the power supply. This improves the output response time for applications that do not require a
physical output disconnect during normal output on/off operation. This parameter is saved in non-
volatile memory.
Parameter Typical Return
0|OFF|1|ON 0๎œœor๎œœ1
Locks the SDS relays closed: OUTP:REL:LOCK ON