Features and Functions 3
Keysight 34980A Mainframe User’s Guide 135
Temperature Measurement Configuration
This section contains information to help you configure the instrument for making
temperature measurements. The table below shows the thermocouple, RTD, and
thermistor types for which the instrument supports direct measurements.
[a] Using ITS-90 software conversions
[b] Using ITPS-68 software conversions
Measurement Units
The instrument can report temperature measurements in °C (Celsius), °F (Fahrenheit),
or K (Kelvins). You can mix temperature units on different channels within the
instrument and on the same module.
–The CONFigure and MEASure? commands automatically select °C.
– Setting the Mx+B measurement label to °C, °F, or K has no effect on the
temperature measurement units currently selected.
– The instrument selects Celsius when the probe type is changed and after a
Factory Reset (*RST command). An Instrument Preset (SYSTem:PRESet
command) or Card Reset (SYSTem:CPON command) does not change the units
setting.
Front Panel Operation:
DMM or Channel (Configure) > TEMPERATURE >
UNITS
Remote Interface Operation:
UNIT:TEMP {C|F|K}[,(@<ch_list>)]
Thermocouple Types
[a]
RTD Types Thermistor Types
B, E, J, K, N, R, S, T R
0
= 49Ω to 2.1 kΩ
α = 0.00385 (DIN/IEC 751)
[a]
α = 0.00391
[b]
2.2 kΩ, 5 kΩ, 10 kΩ
(YSI 44000 Series)