Features and Functions 3
Keysight 34980A Mainframe User’s Guide 161
–The CONFigure and MEASure? commands automatically set the sample count
to 1.
The instrument sets the sample count to 1 after a Factory Reset (*RST
command). An Instrument Preset (SYSTem:PRESet command) or Card Reset
(SYSTem:CPON command) does not change the setting.
Front Panel Operation:
Scan (Configure) > SAMPLE COUNT
Remote Interface Operation:
SAMPle:COUNt
Channel Delay
You can control the pacing of a scan sweep by inserting a delay between
multiplexer channels in the scan list (useful for high-impedance or
high-capacitance circuits). The delay is inserted between the relay closure and the
actual measurement on the channel, in addition to any delay that will implicitly
occur due to relay settling time. The programmed channel delay overrides the
default channel delay that the instrument automatically adds to each channel.