3 Features and Functions
166 Keysight 34980A Mainframe User’s Guide
Non-Sequential Scanning
By default, the instrument scans the list of channels in ascending order from slot 1
through slot 8 (channels are reordered as needed). If your application requires
non-ordered scanning of the channels in the present scan list, you can use the
non-sequential scanning mode.
This feature is available from the remote interface only.
The scanning mode applies to the entire mainframe and cannot be selectively
used on individual modules.
When sequential scanning is enabled (default), the channels in the scan list
are placed in ascending order from slot 1 through slot 8. Duplicate channels
are not allowed. For example, (@2001,1003,1001,1003) will be interpreted as
(@1001,1003,2001).
When sequential scanning is disabled (OFF), the channels remain in the order
presented in the scan list (see exception below). Multiple occurrences of the
same channel are allowed. For example, (@2001,2001,2001) and
(@3010,1003,1001,1005) are valid and the channels will be scanned in the
order presented.
When you specify a range of channels in the scan list, the channels are always
sorted in ascending order, regardless of the scan order setting. Therefore,
(@1009:1001) will always be interpreted as 1001, 1002, 1003, etc.
If you define a scan list with the sequential mode enabled and later disable the
mode, the scan list will not be reordered; however, the scan list will be treated
as a non-sequential list thereafter.
If you have defined a scan list with the sequential mode disabled (OFF) and
later enable the mode, the channels will be reordered.
Non-sequential scan lists are not stored as part of the instrument state by the
*SAV command; in this case, the ordered mode will be enabled and the scan
list will be empty when the instrument state is restored (*RCL command).
The scan order setting is stored in volatile memory and the ordered mode will
be enabled when power is turned off or after a Factory Reset (*RST command).
Remote Interface Operation:
ROUTe:SCAN:ORDered {OFF|ON}