3 Features and Functions
192 Keysight 34980A Mainframe User’s Guide
Keysight TECHNOLOGIES,<Model Number>,<Serial Number>,<Firmware Rev>
A 10-digit string is returned for the Serial Number field. The Firmware Revision
has the form R.RR and indicates the revision of firmware currently in use on the
specified module.
Product Firmware Updates
As new product features and enhancements become available, you can easily
update your mainframe and plug-in module firmware to ensure optimum
compatibility. The latest firmware updates are available from the Keysight 34980A
product page at www.keysight.com/find/34980AUpdates.
Front Panel Operation:
Utility > FIRMWARE > UPDATE
Once you have downloaded the latest mainframe firmware (see above), use the
knob to scroll through the installed modules that require a firmware update. To
exit the menu without installing the updates, select CANCEL.
Instrument State Storage
The instrument has five storage locations in non-volatile memory to store
instrument states, numbered 1 through 5. You can assign a user-defined name to
each of locations 1 through 5.
You can store the instrument state in any of the five locations, but you can only
recall a state from a location that contains a previously stored state.
The instrument stores the state of all plug-in modules including all channel
configurations, scanning setups, and Mx+B scaling values. However, note that
only the measurement attributes of the currently-selected function (range,
resolution, etc.) will be preserved in the stored states.
Before recalling a stored state, the instrument verifies that the same plug-in
module types are installed in each slot. If a different module type is installed,
the instrument will perform the equivalent of a Factory Reset (*RST command)
and an error will be generated.
When shipped from the factory, storage locations 1 through 5 are empty. In
addition, the automatic recall mode is disabled (MEMory:STATe:RECall:AUTO
OFF command) and a Factory Reset (*RST command) is issued when power is
turned on.