Features and Functions 3
Keysight 34980A Mainframe User’s Guide 163
Front Panel Operation:
Channel (Configure) > CHANNEL DELAY > TIME
Once you have added the specified channel to the scan list, the channel delay
choice will be visible in the menu.
Interface Operation:
The following command add a 2-second channel delay to
the specified channels.
ROUT:CHAN:DELAY 2,(@1003,1013)
Automatic Channel Delays
If you do not specify a channel delay, the instrument selects a delay for you. The
delay is determined by the delay based on function, range, integration time, and
ac filter setting.
DC Voltage, Thermocouple, DC Current (for all ranges):
Resistance, RTD, Thermistor (2- and 4-wire):
Integration Time Channel Delay
PLC > 1
PLC 1
2.0 ms
1.0 ms
Range
Ch a n n e l D e l ay
(for PLC > 1)
Channel Delay
(for PLC 1)
100Ω
1 kΩ
10 kΩ
100 kΩ
1 MΩ
10 MΩ
100 MΩ
2.0 ms
2.0 ms
2.0 ms
25 ms
30 ms
200 ms
200 ms
1.0 ms
1.0 ms
1.0 ms
20 ms
25 ms
200 ms
200 ms