Features and Functions 3
Keysight 34980A Mainframe User’s Guide 165
Relative Format (Default):
Absolute Format:
The reading format applies to all readings being removed from the instrument
from a scan; you cannot set the format on a per-channel basis.
–The CONFigure and MEASure? commands automatically turn off the units,
time, channel, and alarm information.
The format settings are stored in volatile memory and will be lost when power
is turned off or after a Factory Reset (*RST command).
Remote Interface Operation:
Use the following commands to select the
reading format.
FORMat:READing:ALARm ON
FORMat:READing:CHANnel ON
FORMat:READing:TIME ON
FORMat:READing:TIME:TYPE {ABSolute|RELative}
FORMat:READing:UNIT ON
1
2
Reading with units (26.195 °C)
Time since start of scan (17 ms)
3
4
Channel number
Alarm limit threshold crossed
(0 = No Alarm, 1 = LO, 2 = HI)
2.61950000E+01 C,000000000.017,1003,2
1 2 4
3
1
2
3
Reading with units (26.195 °C)
Date (November 21, 2004)
Time of day (3:30:23.000 PM)
4
5
Channel number
Alarm limit threshold crossed
(0 = No Alarm, 1 = LO, 2 = HI)
2.61950000E+01 C,2004,11,21,15,30,23.000,1003,2
1 3 542