Features and Functions 3
Keysight 34980A Mainframe User’s Guide 147
Front Panel Operation:
DMM or Channel (Configure) > SCALING > GAIN|OFFSET|UNITS
To define the label on the selected channel, press the arrow keys to move the
cursor to a specific position and then turn the knob to select the desired letter or
number. To clear the label on the selected channel, change each character to “ ^
(starting with the rightmost character) and then press the left arrow key to move
to the next character.
Remote Interface Operation:
Use the following commands to set the gain,
offset, and custom measurement label.
CALC:SCALE:GAIN 1.2,(@1003)
CALC:SCALE:OFFSET 10,(@1003)
CALC:SCALE:UNIT 'PSI',(@1003)
After setting the gain and offset values, send the following command to enable
the scaling function on the specified channel.
CALC:SCALE:STATE ON,(@1003)