1 Introduction to the 34980A
30 Keysight 34980A Mainframe User’s Guide
The figure below shows the external connections required to synchronize the scan
sequence between the 34980A and an external instrument. The 34980A must
notify the external instrument when a relay is closed and fully settled (including
channel delay). The 34980A outputs a Channel Closed pulse. In response, the
external instrument must notify the 34980A when it has finished its measurement
and is ready to advance to the next channel in the scan list. For more information,
see “Scanning With External Instruments” on page 171.