14 Connect the power sensor to the output port of the analyzer. Adjust the
analyzer power output level to produce a reading of 0.0 dBm on the power
meter.
15 Disconnect the power sensor from the output port of the analyzer. Connect the
attenuator to the output port of the analyzer.
16 Connect the probe, the probe adapter, and a barrel between the attenuator
and the power sensor. The signal from the analyzer should now go through the
attenuator and then the probe before reaching the power sensor. Record the
power reading on the “Test Record” on page 47 (item F).
17 Remove the attenuator pad, connect the probe adapter to the output port of
the analyzer. Connect the attenuator pad between the probe and the power
sensor. The signal from the analyzer should now go through the probe and
then the pad before reaching the power sensor. Record the power reading on
the “Test Record” on page 47 (item G).
18 Perform the calculation indicated on the test record to determine the result of
the test.
0.0 dBm is used to allow for possible inaccuracy of power meters