4Operation
34 Keysight 85024A User’s and Service Guide
Operator’s Check
The operator’s check is designed to be a simple functional test for the probe. If the
probe fails the operator’s check, or if you need to verify that the probe meets its
warranted specifications, you will need to do the performance tests as described
in Chapter 5, "Performance Tests".
The operator’s check can be performed with either a network analyzer or a
spectrum analyzer. Use one of the following two procedures to perform the
operator’s check.
Operator’s check using a network analyzer
Required equipment
Procedure
1 Preset the analyzer using the [PRESET] hardkey.
2 Configure the analyzer to measure transmission.
3 Set the output power level of the analyzer to 0 dBm.
4 Set the frequency sweep range on the analyzer from 300 kHz to 3 GHz.
5 Connect a type-N cable (a through cable) between the output and the input
ports on the analyzer.
6 Perform a response calibration on the analyzer. The trace on the analyzer
should now be flat at 0 dB.
Item Part number
Network analyzer Any compatible with type-N connectors
Type-N cable 50 W, male connectors
Probe adapter 11880-60001
NOTE
If you are not familiar with network analyzer operation, refer to “Network
Analyzer Operation” on page 39 for basic information about performing the
analyzer operations used in this procedure.