Service 7
Keysight 85024A User’s and Service Guide 69
Replacing the type-N output connector
Tools required:
9/16 inch open-end wrench
Refer to Figure 7-9.
1 Remove the connector body with the open-end wrench.
2 Remove the male center conductor assembly from the gold plated inner
connector. Use your fingers to pull out the assembly. The small insulator may
or may not come out attached to the center conductor assembly. If it does,
remove it from the old center conductor assembly and place it on the new one.
If it is not attached to the center conductor assembly (it remained inside the
inner connector), proceed to step 3.
3 Hold the new center conductor assembly by the large round insulator, not by
the gold plated center conductor. Carefully align the new center conductor
and insert it into the inner connector. Press gently until you feel it click into
place.
4 Place the new connector body and tighten with the open-end wrench.
Figure 7-9 Output connector parts
NOTE
Refer to Chapter 6, “Replaceable Parts” on page 49 for part numbers.