5 Performance Tests
38 Keysight 85024A User’s and Service Guide
Recommended Test Equipment
Table 5-1 lists the equipment that is recommended for use in performance testing
of the probe. Other equipment may be substituted if its specifications meet or
exceed the specifications listed in the “Critical Specifications” column.
Table 5-1 Recommended test equipment
Type Critical specifications
Recommended
model or part number
Vector network analyzer 300 kHz to 3 GHz
8714ET/ES, 8753ET/ES, 8753E
[a]
[a] Many models of network analyzers can be used for these performance tests. The network analyzer must cover the
frequency range 300 kHz to 3 GHz, and be capable of power output in the range of –3 to +5 dBm.
Power meter 437B/438A or E4418A/E4419A
Power sensor
300 kHz to 3 GHz
–30 dBm to +5 dBm
8482A
Type-N cable, male-to-male 50 W 8120-8862 or equivalent
Probe adapter Supplied with probe 11880-60001
Attenuator, type-N, 50 W Any value 10 to 20 dB nominal
8491A/B/C Option 20, Option 10,
or equivalent
Type-N Adapter,
female-to-female
1250-0777 or equivalent