E8257D/67D, E8663D PSG Signal Generators Service Guide
Assembly Replacement
A2 Display
3-14
A2 Display
Tools Required
• T–10 driver
Removal Procedure
Refer to Figure 3- 7 for instruments with serial number prefix < US/MY5XXX3XXX for this procedure.
Refer to Figure 3- 8 for instruments with serial number prefix > US/MY5XXX3XXX for this procedure.
1. Disconnect the power cord.
2. Remove the front panel. Refer to “Front Panel” on page 3- 8.
3. For serial number prefix < US/MY5XXX3XXX, disconnect the A2DS1W1 cable from the A4 Inverter.
For serial number prefix > US/MY5XXX3XXX, disconnect the A2W1 cable from the A2 Display.
4. Disconnect the W8 ribbon cable from the A3 Power Switch by lifting the white tabs on the connector.
5. Using the T–10 driver, remove the four screws (12) from the A3 Power Switch.
6. Remove the A3 Power Switch.
7. Using the T–10 driver, remove the ten screws (11) that secure the RFI shielding (10) to the front panel.
8. Remove the metal shielding.
9. Using the T–10 driver, remove the four screws (9) that secure the A2 Display to the front panel.
10. Remove the A2 Display.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Torque all T–10 screws to 9 in–lbs.
3. Perform the post–repair adjustments and performance tests that pertain to this removal procedure.