E8257D/67D, E8663D PSG Signal Generators Service Guide
Post–Repair Procedures
CPU Configuration
4-14
CPU Configuration
1. Power on the signal generator.
If the firmware loaded on the CPU matches the instrument model, go to Step 6.
If the firmware is for a different instrument model, the instrument should still turn–on, but the firmware needs
to be changed. To change the firmware continue with Step 2.
2. Set the IP address:
Press
Utility > GPIB/RS–232 LAN > LAN Setup > IP Address.
a. Set the LAN configuration to DHCP.
b. Cycle the signal generator power.
3. Set the remote language:
•Press
Return > Remote Language > SCPI.
4. Check that the instrument identification number (IDN) is correct by executing the command, *IDN?.
If the IDN is not correct, change the IDN by executing the following command:
:SYST:IDN "Agilent Technologies, <Model Number>, <Serial Number>, <Firmware Revision>"
NOTE When executing the command, ensure that the firmware revision parameter matches the signal generator
model. For example, E8257D and E8267D use firmware revisions that start with C.04.xx.
5. Download the new firmware:
a. Go to www/agilent.com/find/psg/upgradeassistant
b. Download (save to disk) both the PSG–ESG Upgrade Assistant software and the latest PSG firmware.
c. Go to the folder where the files were saved and click on each program.
The programs open the files and saves the routines to a folder.
d. Click
Yes when asked to create a short–cut on your desktop.
6. Install the new firmware:
a. Open the PSG–ESG Upgrade Assistant.
Non–Harmonic Spurious
Single- Sideband Phase Noise (Option UNX or UNY test
procedure)
Single- Sideband Phase Noise (Non-Option UNX or UNY test
procedure)
Residual Phase Noise (E8257D/E8267D/E8663D Option UNX or
UNY and without H5Y Only) (Manual Performance Test)
✓✓
Internal EVM Digital Modulation Quality (E8267D Option 601 or
602 Only)
Digital Modulation Power Relative to CW (E8267D Only)
✓✓
Table 4-12 Performance Tests: Assemblies A30—A46, AT1, B1, and RF Output Connector (For prefixes >= xx4928) (Continued)
Performance Tests Replaced Assembly
The following performance tests are listed in the order that
they should be performed to minimize changes in test
equipment configurations.
A30 Mod Filter with High Power
A30 Mod Filter with Standard Power
A31 Motherboard
A32 10 MHz Crystal Oscillator
A34 Internal Hard Drive
A35 3 20 IQ Modulator
A36 Quadraplier
A37 Up Converter
A38 Lowband Switch Filter
A39 Directional Coupler
A40 Compact Flash Drive
A43 LB Amp/Filter
A44 Scan Mod
A45 Frac- N/Offset
A46 Offset Loop
AT1 115 dB Attenuator
B1 Fan
RF Output Connector