EasyManua.ls Logo

Keysight E8257D/67D - A14 Baseband Generator (E8267 D Only)

Keysight E8257D/67D
344 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E8257D/67D, E8663D PSG Signal Generators Service Guide
Assembly Replacement
A14 Baseband Generator (E8267D Only)
3-36
A14 Baseband Generator (E8267D Only)
Tools Required
T–10 driver
needle–nose pliers
Removal Procedure
Refer to Figure 3- 20 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to “Outer Instrument Cover” on page 3- 4 and “Inner
Instrument Cover” on page 3- 6.
3. Position the signal generator with the front panel facing you.
4. Using the T–10 driver, remove the four screws from the Baseband Generator cover.
5. Using the needle–nose pliers, disconnect the W65 flexible cable from A14J800.
6. Disconnect the W64 ribbon cable from A14J13.
7. Disconnect the W90 ribbon cable from A14J21 (Option 602 only).
8. Lift the retention levers and disconnect the A14 Baseband Generator from the A31 Motherboard.
9. Lift the A14 Baseband Generator out of the slot.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post–repair adjustments and performance tests that pertain to this removal procedure.

Table of Contents