Assembly Replacement
A5 Sampler, A6 Frac–N, A7 Reference (Standard and Option UNR/UNX/UNY), A8 Output, A45 Frac-N and A46 Offset
3-23
A5 Sampler, A6 Frac–N, A7 Reference (Standard and Option UNR/UNX/UNY), A8 Output,
A45 Frac-N and A46 Offset
Tools Required
none
Removal Procedure
Refer to Figure 3- 13 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to “Outer Instrument Cover” on page 3- 4 and “Inner
Instrument Cover” on page 3- 6.
3. Position the signal generator with the front panel facing you.
4. Lift the retention levers and disconnect the board assembly from the A31 Motherboard.
5. Lift the board assembly out of the slot.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post–repair adjustments and performance tests that pertain to this removal procedure.
Figure 3-13 A5 Sampler, A6 Frac–N, A7 Reference, A8 Output, A45 Frac-N and A46 Offset
A7
A8
A6/A45
A5/A46